Unit 2: Microscopy
General Microbiology notes · PTU syllabus (BMLS102-18)
On this page
Unit summary
Microscopes make microbes visible. This unit covers the history and types of microscope — light, dark-ground illumination, fluorescence, phase contrast and electron (transmission and scanning) — with their principles and working.
After this unit you can
- Trace the history of microscopy
- Explain magnification and resolving power
- Explain the principles of light, dark-ground, fluorescence and phase contrast microscopes
- Compare transmission and scanning electron microscopes
PTU syllabus topics
- History and types of microscope — light
- DGI
- fluorescent
- phase contrast
- electron (transmission and scanning)
- principles and operational mechanisms of each
Topic 1
History of the microscope
- Janssen (around 1590) made an early compound microscope; Robert Hooke (1665) described cells in cork; Leeuwenhoek made powerful single-lens microscopes (about 270×); Ernst Abbe improved lenses and condensers; Ruska and Knoll built the electron microscope (1931).
Example
Key milestones: Antonie van Leeuwenhoek (1670s) used single-lens microscopes to see 'animalcules' — the first bacteria and protozoa; Ernst Abbe improved lenses and condensers in the 1870s; the electron microscope (Knoll and Ruska, 1931) revealed viruses.
Exam tip
Leeuwenhoek is called the father of microbiology.
Topic 2
Magnification and resolving power
Resolving power = 0.61 λ ÷ NA: shorter wavelengths (electrons) and larger numerical aperture resolve finer detail
Total magnification
Objective × eyepiece (e.g., 100 × 10 = 1,000×)
Resolving power (limit of resolution)
d = 0.61 λ ÷ NA
Numerical aperture
NA = n sin θ; oil (n = 1.515) raises NA
Example
Total magnification: a 10× eyepiece with the 100× oil-immersion objective gives 10 × 100 = 1,000×; with the 40× objective it gives 400×. Resolving power: with green light (λ = 0.55 µm) and an oil objective with NA = 1.4, d = 0.61 × 0.55 ÷ 1.4 ≈ 0.24 µm — two points closer than this cannot be seen separately, however high the magnification. This is why bacteria (about 1 µm) are seen with the oil objective but viruses are not.
- Light microscopes resolve about 0.2 µm; electron microscopes about 0.2 nm. Immersion oil has the same refractive index as glass, preventing light loss at the 100× objective.
Topic 3
Bright-field (light) microscope
Total magnification = objective × eyepiece. Resolution depends on the wavelength of light and the numerical aperture (NA); immersion oil raises NA for the 100× objective.
- Eyepiece (ocular)
- Usually 10×
- Objectives
- 10×, 40×, 100× oil immersion
- Condenser and iris diaphragm
- Focus and control light
- Stage and mechanical stage
- Hold and move the slide
- Coarse and fine adjustment
- Focusing
- Light source
- Built-in lamp or mirror
- Principle: light passes through a stained specimen; objects appear dark on a bright background. Used for Gram and other stained smears.
Example
Calculating magnification: with a 10× eyepiece, the 100× oil-immersion objective gives 1,000× total magnification. The best resolution of a light microscope is about 0.2 µm, so bacteria (about 1–2 µm) are visible but viruses are not.
Exam tip
Resolving power: d = 0.61 λ ÷ NA.
Topic 4
Dark-ground (dark-field) illumination
- A special condenser with a central stop lets only oblique light hit the specimen; scattered light enters the objective, so organisms shine bright on a dark background. Used for unstained, thin organisms such as Treponema pallidum and Leptospira and for motility.
Example
Which microscope for which specimen? Gram-stained sputum smear → bright-field. Fresh genital specimen for the thin spiral Treponema pallidum → dark-field (unstained, bright on a dark background). Sputum with auramine–rhodamine stain for TB → fluorescence. Living, unstained cells in culture, showing internal structures → phase-contrast. Surface of a virus or bacterial cell → electron microscope (SEM) or internal structure (TEM).
Topic 5
Fluorescence microscope
- Specimens stained with fluorochromes absorb short-wavelength (UV or blue) light and emit longer-wavelength visible light. An exciter filter, dichroic mirror and barrier filter separate the light.
- Uses: auramine-rhodamine stain for tuberculosis, immunofluorescence for antigens and antibodies (FTA-ABS, rabies).
Example
Detecting TB: sputum stained with auramine-rhodamine shows Mycobacterium tuberculosis as bright yellow-green rods against a dark background under the fluorescence microscope, allowing faster screening than ZN staining at high magnification.
Exam tip
Fluorescent antibody techniques (direct and indirect) are also used for diagnosis.
Topic 6
Phase contrast microscope
- Converts small differences in refractive index (phase shifts) into differences in brightness using an annular diaphragm and a phase plate, so living, unstained cells and internal structures become visible — useful for motility, cell division and parasites.
Example
Viewing living cells: unstained, motile bacteria or trophozoites in a wet mount are almost invisible in bright-field but appear dark against a bright background under phase contrast, without killing them by staining.
Exam tip
Phase contrast is used to observe live, unstained cells and their movement.
Topic 7
Electron microscopes
Beam
Passes through ultra-thin sections
Scans the specimen surface
Image
2D internal detail
3D surface view
Resolution
About 0.1–0.2 nm
About 1–10 nm
Preparation
Fixation, embedding, ultra-thin sectioning, heavy-metal staining
Coating with gold
Use
Viral structure, organelles
Surface structures
- Principle: a beam of electrons (very short wavelength) is focused by electromagnetic lenses in a vacuum; images appear on a fluorescent screen or detector. Specimens must be dead and dry.
Source
Visible light
Electron beam
Lenses
Glass
Electromagnetic
Magnification
Up to about 1,500×
Up to 1,000,000× or more
Specimen
Living or dead
Dead, in vacuum
Cost
Low
Very high
Example
Choosing an EM: to see the internal structure of a virus particle, a thin section is examined with TEM; to see the surface of a bacterial cell, SEM gives a 3D-looking image. Both need a vacuum, so living cells cannot be viewed.
Exam tip
Electron microscopes resolve down to about 0.1–0.2 nm, far below the light microscope's 0.2 µm.
Key terms
- Resolving power
- Ability to distinguish two close points as separate
- Numerical aperture
- Measure of a lens's light-gathering ability
- Dark-ground microscopy
- Bright objects on a dark background
- Fluorochrome
- Dye that fluoresces under UV light
- TEM
- Microscope transmitting electrons through thin sections
Quick revision
- History: Janssen, Hooke, Leeuwenhoek, Abbe, Ruska.
- Magnification; resolution 0.61 λ ÷ NA; oil immersion.
- Bright-field, dark-ground, fluorescence, phase contrast.
- TEM vs SEM; light vs electron microscope.
Important exam questions
Practice questions written to the PTU exam pattern for this unit's syllabus: short answers (Section A style) and long answers (Sections B and C style).
Short-answer questions
- Q1.Why is oil used with the 100× objective?
- Q2.Define resolving power.
- Q3.Which microscope is used to see Treponema pallidum?
- Q4.What is a fluorochrome?
- Q5.What is the advantage of phase contrast microscopy?
- Q6.Distinguish TEM and SEM.
Long-answer questions
- Q1.Describe the parts and working of a compound microscope.
- Q2.Explain the principles of dark-ground and fluorescence microscopy.
- Q3.Explain phase contrast microscopy.
- Q4.Compare light and electron microscopes.
Stuck on this unit?
Message SBS on WhatsApp for help with General Microbiology, or to ask about studying B.Sc MLS at Synetic.
